Evaluation and deterioration test of SOFC flat plate single cell
BEL-SOFC is the automatic evaluation and deterioration test system for single plate cell in SOFC. Unique sample setting mechanism enables highly reproducible measurement.
Gas composition, gas flow and temperature can be controlled by dedicated software. Combining with electrochemical instruments, impedance, OCP and I-V properties can be measured.
Standard cell (Ni-YSZ |YSZ |LSM dia.:approx.
20mm) is prepared. It is useful to investigate the various performances and to compare with customer's cell.
| Sample | Single flat plate(dia.:approx. *Other diameters are negotiable |
|---|---|
| Gases | Anode Mixture of H2, N2,CH4M,CO and saturated water vapor Cathode Mixture of O2 and N2 |
| Gas flow control | Mass flow controller x 2 (F.S. : 200sccm, Accuracy +/- 2%) |
| Humidity Control | Bubbling method(5-80oC) |
| Temperature Control | Electric furnace:400-1000oC(accuracy: Thermostatic air oven:40-100oC(accuracy: Bubbling + BEL-thermo:5-80oC(accuracy: Flexible heater tube:50-100oC(accuracy: |
| Pressure | 100kPa (Abs) |
| Electrochemical evaluation method | Impedance measurement and I-V measurement by 3 terminal (Each electrode case) and 4 terminal pole (overall cell case) |
| Condenser | Set 15oC by circulation water bath |
| Safety measure | Short circuit breaker, lowering water level alarm, hydrogen detection alarm, CO detection alarm, Methane detection alarm Temperature highest limit alarm(heater OFF), Heater disconnection alarm (only for TIC1) |
| Measurement software | Condition setting:Measurement sequence setting (Temp.・flow・time), Synchronism between the sequence setting and impedance measurement and I-V measurement. Saving format:CSV file Display:Flow chart/TIC1-4, Trend graph of MFC input value/I-V |
| Physical | W1100xD600xH1800mm(body:W750xD600xH1425mm) |